EasyManua.ls Logo

ITech IT-M3100 - [Source:]Voltage:slew[:Both] <Nrf+>,<Nrf; [Source:]Voltage:slew:negative <Nrf

ITech IT-M3100
114 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Source Commands
Copyright © Itech Electronic Co., Ltd. 26
Arguments
<NRf+> MIN TO MAX
Default value
MIN
Example
VOLT:TRIG 10.00
Query command
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]? [MINimum|MAXimum]
Returns
NR3
[SOURce:]VOLTage:SLEW[:BOTH] <NRf+>,<NRf+>
This command sets the voltage rising and falling slew rate of the power supply.
Syntax
[SOURce:]VOLTage:SLEW[:BOTH] <NRf+>,<NRf+>
Arguments
<0.025-9.999>,<0.025-9.999>
Default value
0.025,0.100
Example
VOLT:SLEW 0.03,0.03
Query command
[SOURce:]VOLTage:SLEW[:BOTH]?
[MINimum|MAXimum],[MINimum|MAXimum]
Returns
<NR3>,<NR3>
[SOURce:]VOLTage:SLEW:NEGative <NRf+>
This command sets the voltage falling slew rate of the power supply.
Syntax
[SOURce:]VOLTage:SLEW:NEGative <NRf+>

Table of Contents