EasyManua.ls Logo

K-Patents PR-23 - PR-23-GC Specifications

K-Patents PR-23
224 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
100
PR-23 instrucon manual
9.5.2 PR-23-GC specicaons
General speciications
Refractive Index range: Full range n
D
1.3200–1.5300 (corresponds to
0–100 % b.w.), Sapphire prism
Accuracy: Refractive index n
D
± 0.0002 (corresponds typ-
ically to ± 0.1 % by weight)
Repeatability and stability correspond to accu-
racy
Speed of response: 1 s undamped, damping time selectable up to
5 min
Calibration: With Cargille certiied refractive index liquids
over full range of n
D
1.3200–1.5300
CORE-Optics: No mechanical adjustments (US Patent No. US6067151)
Digital measurement: 3648 pixel CCD element
Light source: Light emitting diode (LED) 589 nm wavelength,
sodium light
Temperature sensor: Built-in Pt-1000
Temperature compensation: Automatic, digital compensation
Instrument veriication: With certiied refractive index liquids and
K-Patents documented procedure
Ambient temperature: Sensor: max. 45 °C (113 °F),
min. -20 °C (-4 °F)
Indicating transmitter: max. 50 °C (122 °F),
min. 0 °C (32 °F)
SENSOR PR-23-GC:
Process connection by Sandvik coupling L 76.1 mm (2.5 inch) for
(in upper elbow of pipe): pipe line sizes of 2.5 inch and larger; via reduc-
ing ferrule PR-9283 for 2 inch pipes
Wafer low cell WFC connection via Wafer low cell WFC for pipe line sizes 15 mm
(in straight pipe): (0.5 inch), 25 mm (1 inch) and 40 mm (1.5 inch);
Wafer low cell body mounts between ANSI 150 psi,
DIN PN 25 or JIS
Pipe low cell PFC connection via a Pipe low cell PFC for pipe line sizes 15 mm
(in straight pipe) (0.5 inch), 25 mm (1 inch) and 40 mm (1.5 inch);
Pipe low process connection ANSI 150 psi, DIN
PN 25 or JIS
Process pressure: up to 15 bar (200 psi) at 20 °C (70 °F)
Process temperature: -40 °C–+130 °C (-40 °F–+266 °F)
Process wetted parts, standard: AISI 316L stainless steel, prism sapphire, prism
gaskets PTFE (telon)
Process wetted parts, options: AISI 904L stainless steel, Alloy 20, Hastelloy
C-276, Nickel 200, Titanium ASTM B348 or SAF
2205
Sensor protection class: IP67, Nema 4X
Sensor weight: 2.0 kg (4.4 lbs)

Table of Contents