13 Sensor vericaon
193
If veriication fails even on the repetition of the veriication procedure, ill in the
PR-23 sensor veriication form (available in Appendix C and send it to K-Pa-
tents or your nearest K-Patents representative or email the collected information to
<info@kpatents.com> and wait for further instructions.
For the sensor veriication form, you need to collect data from the Veriication step 2
display and the Veriication results display. The sensor’s serial number is shown in
the upper right corner of each display. The measured n
D
(RI) is given when VERIFY has
been pressed in Veriication step 2. The list of CCD and TEMP values are found on the
veriication results display, see (Figure 13.10)
Figure 13.10 Finding vericaon informaon for PR-23 sensor vericaon form