IEEE-488 Reference
3-42
This command is used to configure the instrument for subsequent measurements on the specified
function. Basically, this command places the instrument in a “one-shot” measurement mode.
The :READ? command can then be used to trigger a measurement and acquire a reading (see
:READ?).
When this command is sent, the Model 2002 will be configured as follows:
The function specified by this command is selected.
All controls related to the selected function are defaulted to the *RST values.
Continuous initiation is disabled (:INITiate:CONTinuous OFF).
The control sources in all three layers of the Trigger Model are set to Immediate.
The count values in all three layers of the Trigger Model are set to one (1).
The delay in all three layers of the Trigger Model are set to zero (0).
The Model 2002 is placed in the idle state.
All math calculations are disabled.
Buffer operation is disabled. A storage operation currently in process will be aborted.
Autozero controls are set to the *RST default values.
The acquisition method is set to normal.
All operations associated with switching cards (scanning) are disabled.
This command is automatically asserted when the :MEASure? command is sent.
PRINT #1, "output 16; :conf:volt:dc" ' Perform :CONFigure operations.
:READ?
This command is typically used to acquire a post-processed reading after the instrument has
been configured by the :CONFigure command. The :CONFigure command places the instru-
ment in a “one-shot” measurement mode. Every time the :READ? command is then sent, a mea-
surement conversion will occur and the reading will be acquired.
When this command is sent, the following commands will be executed in the order that they are
presented:
:ABORt
:INITiate
:FETCh?
When :ABORt is executed, the instrument will go into the idle state if continuous initiation is
disabled. Otherwise, operation will re-start at the beginning of the Trigger Model.
If the instrument is in the idle state, :INITiate will take the instrument out of the idle state. If
continuous initiation is enabled, (:INITiate:CONTinuous ON), then the :INITiate command will
generate an error and ignore the command.
The :FETCh? for details. Note that an “Init ignored” error will not cancel the execution of the
:FETCh? command.
PRINT #1, "output 16; :conf:volt:dc" ' Select “one-shot” DCV function
PRINT #1, "output 16; :read?" ' Perform :READ? operations
PRINT #1, "enter 16" ' Get response message from 2002
Description
Program fragment
Description
Program fragment

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A

Related product manuals