EasyManua.ls Logo

Keysight Technologies J-BERT M8020A - Page 407

Keysight Technologies J-BERT M8020A
502 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight M8070A Programming Guide 407
SCPI Command Reference 5
:PLUGin:EDIagram:FETCh[:RESult]:CVOLtage?
Syntax :PLUGin:EDIagram:FETCh[:RESult]:CVOLtage? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
Description This command returns the cross voltage which is the measurement of the
crossing point level in relation to the logic 1 level and logic 0 level.
This SCPI is applicable for M8041A, M8051A and M8062A.
Example :PLUG:EDI:FETC:CVOL? "MyMeasurement"
:PLUGin:EDIagram:FETCh[:RESult]:DCDistortion?
Syntax ::PLUGin:EDIagram:FETCh[:RESult]:DCDistortion? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
Description This command returns the value of DC Distortion calculated after a run.
This value is the difference between the period of a 1 bit and a 0 bit.
This SCPI is applicable for M8041A, M8051A and M8062A.
Example :PLUG:EDI:FETC:DCD? ‘MyMeasurement’
:PLUGin:EDIagram:FETCh[:RESult]:FALLtime?
Syntax :PLUGin:EDIagram:FETCh[:RESult]:FALLtime? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement nameDescription
Description This command returns the value of the fall time. Fall time is a
measurement of the mean transition time of the data on the downward
slope of an eye diagram. This transition time is either 2080 or 1090.
See :PLUGin:EDIagram:EVALuation:TTIMe[?] on page 404.
This SCPI is applicable for M8041A, M8051A and M8062A.
Example :PLUG:EDI:FETC:FALL? ‘MyMeasurement’

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Related product manuals