408 Keysight M8070A Programming Guide
5 SCPI Command Reference
:PLUGin:EDIagram:FETCh[:RESult]:HEIGht?
Syntax :PLUGin:EDIagram:FETCh[:RESult]:HEIGht? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
Description This command returns the value of the eye height which is a measurement
of the vertical opening of an eye diagram. This opening is affected by the
BER threshold that is set by the BER threshold command. See
:PLUGin:EDIagram:EVALuation:BERThresh[?] on page 403
This SCPI is applicable for M8041A, M8051A and M8062A.
Example :PLUG:EDI:FETC:HEIG? ‘MyMeasurement’
:PLUGin:EDIagram:FETCh[:RESult]:HILevel?
Syntax :PLUGin:EDIagram:FETCh[:RESult]:HILevel? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
Description This command returns the measurement of the mean value of the logical 1
in the eye diagram. This is directly affected by the values set for Eye
Boundaries. One Level is a measure of the mean value of the logical 1 of an
eye diagram.
This SCPI is applicable for M8041A, M8051A and M8062A.
Example :PLUG:EDI:FETC:HIL? ‘MyMeasurement’
:PLUGin:EDIagram:FETCh[:RESult]:JPPeak?
Syntax :PLUGin:EDIagram:FETCh[:RESult]:JPPeak? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
Description This command returns Jitter P-P value. It is the full width of the eye
diagram at the eye crossing point. This is affected by the BER threshold
that is set by the BER threshold command. See
:PLUGin:EDIagram:EVALuation:BERThresh:THReshold[?] on page 404
This SCPI is applicable for M8041A, M8051A and M8062A.
Example :PLUG:EDI:FETC:JPP? ‘MyMeasurement’

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Related product manuals