444 Keysight M8070A Programming Guide
5 SCPI Command Reference
This subnode has the following commands:
Table 88
:PLUGin:CCAPture:FETCh:ERRor:COUNt?
Syntax :PLUGin:CCAPture:FETCh:ERRor:COUNt? ‘identifier’ [,‘Location’]
Input
Parameters
‘identifier’: Specify the measurement name.
[,‘Location’]: ‘M*.DataIn1’ or ‘M*.DataIn2’
Description This command gives the error count of the captured data. If simply run
without specifying the location, then it returns the results for the current
run.
This SCPI is applicable for M8041A and M8051A.
Example :PLUGin:CCAPture:FETCh:ERRor:COUNt? ‘MyMeasurement’,‘M1.DataIn1’
Name Description under
:ERRor:COUNt?
:PLUGin:CCAPture:FETCh:ERRor:COUNt?
on page 444
:PATTern:DATA:HOLDoff? :PLUGin:CCAPture:FETCh:PATTern[:DATA
]:HOLDoff?
on page 445
:PATTern:DATA:DEPTh? :PLUGin:CCAPture:FETCh:PATTern[:DATA
]:DEPTh?
on page 445
:PATTern:ERRor:SYMBol:FIRSt? :PLUGin:CCAPture:FETCh:PATTern:ERRor
:SYMBol:FIRSt?
on page 446
:PATTern:ERRor:SYMBol:LAST? :PLUGin:CCAPture:FETCh:PATTern:ERRor
:SYMBol:LAST?
on page 446
:RUN[:STATus]? :PLUGin:CCAPture:FETCh:RUN[:STATus]?
on page 447