M8190A User Interface 2
M8190A User’s Guide 75
Output_SR: The user-defined output sampling rate will be used to
calculate the best value for the output record length being a multiple of
the granularity for the current DAC mode according to the time window of
the input signal. Final time window will be slightly adjusted to keep the
selected output sampling rate. This change is reflected in the Source
Sampling Rate numeric entry field value.
Output_RL: The user-defined output Waveform Length will be used to
calculate the best value for the output Sample Rate according to the time
window of the input signal. Waveform Length will be adjusted to the
nearest multiple of the granularity for the current DAC mode according to
the time window of the input signal.
Zero_Padding: Output Waveform Length is calculated based on the input
waveform time window and the user-defined output sampling rate. The
resulting waveform length will not be, in general, a multiple of the
granularity. To meet the granularity conditions, a number of zero samples
are added until the combined number of samples is a multiple of the
granularity. Output Sample Rate will be slightly adjusted to keep the input
waveform time window.
Truncate: Output Waveform Length is calculated based on the input
waveform time window and the user-defined output sampling rate. The
resulting waveform length will not be, in general, a multiple of the
granularity. To meet the granularity conditions, a number of samples is
removed until the resulting number of samples is a multiple of the
granularity. Output Sample Rate will be slightly adjusted to keep the input
waveform time window.
Repeat: Output Waveform Length is calculated based on the input
waveform time window and the user-defined output sampling rate. The
resulting waveform length will not be, in general, a multiple of the
granularity. To meet the granularity conditions, the base waveform is
repeated the minimum number of times so the overall number of samples
is a multiple of the granularity. Output Sample Rate will be slightly
adjusted to keep the input waveform time window. The Waveform Length
field will show the length of the combined waveform.
Waveform Length
It shows the number of samples of the resampled output waveform. It can be set
when Resampling Mode is Output_RL. Otherwise, this field is an indicator. This field
is not active if File Format is the "SigStudioEncrypted".