Output_SR: The user-defined output sampling rate will be used to 
calculate the best value for the output record length being a multiple of 
the granularity for the current DAC mode according to the time window of 
the input signal. Final time window will be slightly adjusted to keep the 
selected output sampling rate. This change is reflected in the Source 
Sampling Rate numeric entry field value. 
  Output_RL: The user-defined output Waveform Length will be used to 
calculate the best value for the output Sample Rate according to the time 
window of the input signal. Waveform Length will be adjusted to the 
nearest multiple of the granularity for the current DAC mode according to 
the time window of the input signal. 
  Zero_Padding: Output Waveform Length is calculated based on the input 
waveform time window and the user-defined output sampling rate. The 
resulting waveform length will not be, in general, a multiple of the 
granularity. To meet the granularity conditions, a number of zero samples 
are added until the combined number of samples is a multiple of the 
granularity. Output Sample Rate will be slightly adjusted to keep the input 
waveform time window. 
  Truncate: Output Waveform Length is calculated based on the input 
waveform time window and the user-defined output sampling rate. The 
resulting waveform length will not be, in general, a multiple of the 
granularity. To meet the granularity conditions, a number of samples is 
removed until the resulting number of samples is a multiple of the 
granularity. Output Sample Rate will be slightly adjusted to keep the input 
waveform time window. 
  Repeat: Output Waveform Length is calculated based on the input 
waveform time window and the user-defined output sampling rate. The 
resulting waveform length will not be, in general, a multiple of the 
granularity. To meet the granularity conditions, the base waveform is 
repeated the minimum number of times so the overall number of samples 
is a multiple of the granularity. Output Sample Rate will be slightly 
adjusted to keep the input waveform time window. The Waveform Length 
field will show the length of the combined waveform. 
  Waveform Length 
It shows the number of samples of the resampled output waveform. It can be set 
when Resampling Mode is Output_RL. Otherwise, this field is an indicator. This field 
is not active if File Format is the "SigStudioEncrypted".