Pulse
Back to Arb Commands
Arb Type Current Voltage
Pulse
[SOURce:]ARB:CURRent:PULSe:END:TIMe <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:PULSe:END:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:PULSe:END:TIMe<value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:PULSe:END:TIMe? [MIN|MAX,] (@<chanlist>)
Specifies the time in seconds, after the pulse completes, that the starting current or voltage level
persists. Referenced to t2 in the Pulse diagrams.
[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:END:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:END:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:END:TIMe<value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:END:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the time in seconds, after the pulse completes, that the starting current or voltage level
persists for Arb sequence steps.
Parameter Typical Return
0 262.144 | MIN | MAX, *RST 0 <end time>
Step 0 - 99, *RST 0 (none)
Programs an end time for the current pulse: ARB:CURR:PULS:END:TIM 2,(@1)
Programs an end time for the voltagepulse:ARB:VOLT:PULS:END:TIM 2,(@1)
Programs an end time for step 0: ARB:SEQ:STEP:CURR:PULS:END:TIM 2, 0,(@1)
[SOURce:]ARB:CURRent:PULSe:STARt[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:PULSe:STARt[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:PULSe:STARt[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:PULSe:STARt[:LEVel]? [MIN|MAX,] (@<chanlist>)
Specifies the current or voltage before and after the pulse occurs. Referenced to I0 and V0 in the Pulse
diagrams.
Keysight N6705C Operating and Service Guide 241