[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:STARt[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:STARt[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:STARt[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:STARt[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the current or voltage before and after the pulse occurs for Arb sequence steps.
Parameter Typical Return
Current: 0 - maximum|MIN|MAX, *RST MIN
Voltage: 0 - maximum|MIN|MAX, *RST MIN
The maximum value varies, depending on the currentor voltage
rating of the power module.
<start value>
Step 0 - 99, *RST 0 (none)
Programs a start- and end-level current: ARB:CURR:PULS:STAR 1,(@1)
Programs a start- and end-level voltage: ARB:VOLT:PULS:STAR 1,(@1)
Programs a start- and end-level for step 0: ARB:SEQ:STEP:CURR:PULS:STAR 1, 0,(@1)
[SOURce:]ARB:CURRent:PULSe:STARt:TIMe <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:PULSe:STARt:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:PULSe:STARt:TIMe<value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:PULSe:STARt:TIMe? [MIN|MAX,] (@<chanlist>)
Specifies the delay in seconds after the trigger is received, but before the pulse occurs. Referenced to
t0 in the Pulse diagrams.
[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:STARt:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:STARt:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:STARt:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:STARt:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the delay in seconds after the trigger is received, but before the pulse occurs for Arb
sequence steps.
Parameter Typical Return
0 262.144 | MIN | MAX, *RST 0 <start time>
Step 0 - 99, *RST 0 (none)
Programs a start time for the current pulse: ARB:CURR:PULS:STAR:TIM 1,(@1)
Programs a start time for the voltage pulse: ARB:VOLT:PULS:STAR:TIM 1,(@1)
Programs a start time for step 0: ARB:SEQ:STEP:CURR:PULS:STAR:TIM 1, 0,(@1)
7 SCPI Programming Reference
242 Keysight N6705C Operating and Service Guide