[SOURce:]ARB:CURRent:PULSe:TOP[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:PULSe:TOP[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:PULSe:TOP[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:PULSe:TOP[:LEVel]? [MIN|MAX,] (@<chanlist>)
Specifies the top current or voltage level of the pulse. Referenced to I1 and V1 in the Pulse diagrams.
[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:TOP[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:TOP[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:TOP[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:TOP[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the top current or voltage level of the pulse for Arb sequence steps.
Parameter Typical Return
Current: 0 - maximum|MIN|MAX, *RST MIN
Voltage: 0 - maximum|MIN|MAX, *RST MIN
The maximum value varies, depending on the currentor voltage
rating of the power module.
<top value>
Step 0 - 99, *RST 0 (none)
Programs the top level current: ARB:CURR:PULS:TOP 2,(@1)
Programs the top level voltage: ARB:VOLT:PULS:TOP 20,(@1)
Programs the top level for step 0: ARB:SEQ:STEP:CURR:PULS:TOP 2, 0,(@1)
[SOURce:]ARB:CURRent:PULSe:TOP:TIMe <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:PULSe:TOP:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:PULSe:TOP:TIMe<value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:PULSe:TOP:TIMe? [MIN|MAX,] (@<chanlist>)
Specifies the time of the pulse in seconds. Referenced to t1 in the Pulse diagrams.
[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:TOP:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:TOP:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:TOP:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:TOP:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the time of the pulse in seconds for Arb sequence steps.
Parameter Typical Return
0 – 262.144 | MIN | MAX, *RST 0 <end time>
Step 0 - 99, *RST 0 (none)
Programs a top time for the current pulse: ARB:CURR:PULS:TOP:TIM 2,(@1)
Programs a top time for the voltagepulse: ARB:VOLT:PULS:TOP:TIM 2,(@1)
Programs a top time for step 0: ARB:SEQ:STEP:CURR:PULS:TOP:TIM 2, 0,(@1)
Keysight N6705C Operating and Service Guide 243
7 SCPI Programming Reference