:MAX
:DOFFset? Returns the delay offset required for this instrument.
:DELay
:FALL <value>, (@chanlist) Sets the output turn-off sequence delay.
:RISE <value>, (@chanlist) Sets the output turn-on sequence delay.
:PMODE VOLTage|CURRent, (@chanlist) Sets the mode for turn-on/off transitions. (only on N6761A, N6762A)
:TMODE HIGHz|LOWZ, (@chanlist) Specifies the turn-on/turn-off impedance. (only on N678xA SMU)
:INHibit
:MODE LATChing|LIVE|OFF Sets the operating mode of the remote inhibit digital pin.
:PON
:STATe RST|RCL0 Set the output power-on state.
:PROTection
:CLEar (@chanlist) Resets the latched protection.
:COUPle0|OFF|1|ON Enables/disables channel coupling for protection faults.
:DELay <value>, (@chanlist) Sets the over-current protection programming delay.
:OSCillation0|OFF|1|ON, (@chanlist) Enables/disables output oscillation protection. (only on N678xA SMU)
:TEMPerature
:MARGin? (@chanlist) Returns the margin remaining before the over-temperature trips.
:WDOG
[:STATe] 0|OFF|1|ON Enables/disables the I/O watchdog timer.
:DELay <value> Sets the watchdog delay time.
:RELay
:POLarity NORMal|REVerse, (@chanlist) Sets the polarity of the output relays. (only on Option 760)
SENSe
:CURRent
:CCOMpensatre 0|OFF|1|ON, (@chanlist) Enables/disables the capacitive current compensation.
(does NOT apply to N678xA SMU)
[:DC]
:RANGe
[:UPPer] <value>, (@chanlist) Selects a DC current measurement range.
:AUTO 0|OFF|1|ON, (@chanlist) Enables/disables seamless measurement autoranging.
(only on N6781A/82A/85A/86A)
:DLOG
:CURRent
[:DC]
:RANge
[:UPPer] <value>, (@chanlist) Selects the internal data log current measurement range.
:AUTO 0|OFF|1|ON, (@chanlist) Enables/disables seamless measurement autoranging.
(only on N6781A/82A/85A/86A)
:FUNCtion
:CURRent 0|OFF|1|ON, (@chanlist) Enables/disables current data logging.
:MINMax 0|OFF|1|ON Enables/disables min/max internal data logging.
:VOLTage 0|OFF|1|ON, (@chanlist) Enables/disables voltage data logging.