Parameter Typical Return
(none) <HIGHvalue>
Returns the measured high level currentMEAS:CURR:HIGH? (@1)
Returns the measured high level voltage MEAS:VOLT:HIGH? (@1)
MEASure[:SCALar]:CURRent:LOW? (@<chanlist>)
MEASure[:SCALar]:VOLTage:LOW? (@<chanlist>)
Initiates, triggers, and returns the Low level of a pulse waveform. Values returned are either in
amperes, or volts.See Measurement Data Formats. The reading is in the form +n.nnnnnnE+nn for
each channel specified. Multiple responses are separated by commas.
Parameter Typical Return
(none) <LOW value>
Returns the measured low level currentMEAS:CURR:LOW? (@1)
Returns the measured low level voltage MEAS:VOLT:LOW? (@1)
MEASure[:SCALar]:CURRent:MAXimum? (@<chanlist>)
MEASure[:SCALar]:VOLTage:MAXimum? (@<chanlist>)
MEASure[:SCALar]:CURRent:MINimum? (@<chanlist>)
MEASure[:SCALar]:VOLTage:MINimum? (@<chanlist>)
Initiates, triggers, and returns the maximum or minimum values of a measurement. Values returned
are either in amperes, or volts. The reading is in the form +n.nnnnnnE+nn for each channel specified.
Multiple responses are separated by commas.
Parameter Typical Return
(none) <MIN value>
<MAX value>
Returns the measured minimum current MEAS:CURR:MIN? (@1)
Returns the measured minimum voltage MEAS:VOLT:MIN? (@1)
Returns the measured maximum currentMEAS:CURR:MAX? (@1)
Returns the measured maximum voltage MEAS:VOLT:MAX? (@1)
MEASure:ARRay:CURRent[:DC]? (@<chanlist>)
MEASure:ARRay:VOLTage[:DC]? (@<chanlist>)
MEASure:ARRay:POWer[:DC]? (@<chanlist>)
Initiates and triggers a measurement; returns a list of the digitized output measurement samples.
Values are either in amperes, volts, or watts.
The sampling rate is set by SENSe:SWEep:TINTerval. The position of the trigger relative to the
beginning of the data buffer is determined by SENSe:SWEep:OFFSet:POINts. The number of points
returned is set by SENSe:SWEep:POINts.
7 SCPI Programming Reference
302 Keysight N6705C Operating and Service Guide