4 System
4.6 Alignments
Query Temperature of Last Align Now Receiver
Remote Command
:CALibration:TEMPerature:INTernal:RECeiver?
Example
:CAL:TEMP:INT:REC?
Notes Value in degrees Centigrade when the last successful Align Now Receiver was executed
Dependencies Only appears in VXT models M9410A/11A
State Saved No
Query Time since Last Align Now Source
Remote Command
:CALibration:TIME:INTernal:SOURce?
Example
:CAL:TIME:INT:SOUR?
Notes Value in hours since the last successful Align Now Source
Dependencies Only appears in VXT models M9410A/11A
State Saved No
Query Temperature of Last Align Now Source
Remote Command
:CALibration:TEMPerature:INTernal:SOURce?
Example
:CAL:TEMP:INT:SOUR?
Notes Value in degrees Centigrade when the last successful Align Now Source was executed
Dependencies Only appears in VXT models M9410A/11A
State Saved No
Query Time since Last Align Now Fast
Remote Command
:CALibration:TIME:INTernal:FAST?
Example
:CAL:TIME:INT:FAST?
Notes Value in hours since the last successful Align Now Fast
Dependencies Only appears in VXT models M9410A/11A/15A/16A
State Saved No
Query Temperature of Last Align Now Fast
Remote Command
:CALibration:TEMPerature:INTernal:FAST?
Example
:CAL:TEMP:INT:FAST?
Notes Value in degrees Centigrade when the last successful Align Now Fast was executed
Dependencies Only appears in VXT models M9410A/11A/15A/16A
State Saved No
Vector Modulation Analyzer Mode User's &Programmer's Reference 1926