CALCulate:TRANsform:HISTogram:CLEar[:IMMediate]
Clears the histogram data and restarts histogram ranging if it is enabled (CALCu-
late:TRANsform:HISTogram:RANGe:AUTO ON).
Parameter Typical Return
(none) (none)
Clear the histogram data:
CALC:TRAN:HIST:CLE
l Does not clear measurements in reading memory.
l To clear statistics, limits, histogram data, and measurement data, use CALCulate:CLEar[:IMMediate].
CALCulate:TRANsform:HISTogram:COUNt?
Returns the number of measurements collected since the last time the histogram was cleared.
Parameter Typical
Return
(none) +87
Return the number of measurements used to compute the current histogram:
CALC:TRAN:HIST:COUN?
CALCulate:TRANsform:HISTogram:POINts {<value>|MIN|MAX|DEF}
CALCulate:TRANsform:HISTogram:POINts? [{MIN|MAX|DEF}]
Sets the number of bins between the lower and upper range values for the histogram. Two additional bins
always exist: one for measurements below the lower range and one for measurements above the upper
range.
Parameter Typical Return
{10|20|40|100|200|400|MIN|MAX|DEF}. Default: 100. +100
See
Example
.
l You can specify the lower and upper range values using CALCulate:TRANsform:HISTogram:RANGe:
{LOWer|UPPer}. Lower and upper range values are computed automatically if CALCu-
late:TRANsform:HISTogram:RANGe:AUTO is ON.
l This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
(SYSTem:PRESet).
Keysight Truevolt Series Operating and Service Guide
219
SCPI Programming Reference