[SENSe:]CAPacitance:SECondary {"OFF"|"CALCulate:DATA"}
[SENSe:]CAPacitance:SECondary?
Selects the secondary measurement function for capacitance measurements.
Parameter Typical Return
{"OFF"|"CALCulate:DATA"}, default "OFF" "CALC:DATA"
Configure capacitance and secondary measurements. Make capacitance measurement and retrieve sec-
ondary measurement.
CONF:CAP AUTO,MAX
CAP:SEC "CALC:DATA"
READ?;DATA2?
Typical Response: +3.01566373E-10;+3.01566373E-10
l "CALCulate:DATA" - The measurement value before any math operations are done (including NULL).
l Use READ? or INITiate to start the measurement. Use [SENSe:]DATA2? to retrieve the secondary meas-
urement.
l This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
(SYSTem:PRESet).