[SENSe:]TEMPerature:TRANsducer:TCouple:CHECk {ON|1|OFF|0}
[SENSe:]TEMPerature:TRANsducer:TCouple:CHECk?
Disables or enables the thermocouple check feature to verify that your thermocouples are properly con-
nected for measurements. When enabled, the instrument measures the resistance after each ther-
mocouple measurement to ensure a proper connection. If an open connection is detected (greater than
5kΩ on the 10kΩ range), the instrument reports an overload condition.
Parameter Typical Return
{ON|1|OFF|0}. Default: OFF. 0(OFF)or1(ON)
Configure K-type thermocouple measurements. Select fixed reference junction and set reference junction
temperature to +20°C. Enable thermocouple check. Make and read two measurements:
CONF:TEMP TC,K
TEMP:TRAN:TC:RJUN:TYPE FIX
TEMP:TRAN:TC:RJUN 20.0
TEMP:TRAN:TC:CHEC ON
SAMP:COUN 2
READ?
Typical Response: +3.31785468E+00,+3.61554515E+00
l Applies only to the 34465A and 34470A.
l Enabling the thermocouple check feature slows measurement rates because two measurements have
to be made for each temperature measurement values produced.
l This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
(SYSTem:PRESet).
[SENSe:]TEMPerature:TRANsducer:TCouple:RJUNction {<temperature>|MIN|MAX|DEF}]
[SENSe:]TEMPerature:TRANsducer:TCouple:RJUNction? [{MIN|MAX|DEF}]
Sets the fixed reference junction temperature in degrees Celsius for thermocouple measurements.
Parameter Typical Return
-20°C to +80°C. Default: 0°C. +2.00000000E+01 (always in degrees Celsius).
Configure K-type thermocouple measurements. Select fixed reference junction and set reference junction
temperature to +20°C. Enable thermocouple check. Make and read two measurements:
CONF:TEMP TC,K
TEMP:TRAN:TC:RJUN:TYPE FIX
TEMP:TRAN:TC:RJUN 20.0
TEMP:TRAN:TC:CHEC ON
SAMP:COUN 2
READ?
Typical Response: +3.31785468E+00,+3.61554515E+00
Keysight Truevolt Series Operating and Service Guide
367
SCPI Programming Reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Truevolt Series and is the answer not in the manual?

Keysight Truevolt Series Specifications

General IconGeneral
BrandKeysight
ModelTruevolt Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals