SYSTem:COMMunicate:GPIB:ADDRess <address>
SYSTem:COMMunicate:GPIB:ADDRess?
Assigns instrument's GPIB (IEEE-488) address, which is displayed at power-on. Each device on the GPIB
interface must have a unique address.
The GPIB interface is an optional feature. See Models and Options for details.
Parameter Typical Return
0to30,default22 +15
Set GPIB address to 15:
SYST:COMM:GPIB:ADDR 15
l Your computer's GPIB interface card has its own address. Avoid using this address for any instrument
on the GPIB bus.
l Cycle power for this command to take effect.
l GPIB is enabled or disabled by SYSTem:COMMunicate:ENABle.
l The GPIB interface is an optional feature. See Models and Options for details.
l This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instru-
ment Preset (SYSTem:PRESet).
l This parameter is set to its default value when the instrument is shipped from the factory and after
SYSTem:SECurity:IMMediate.
Keysight Truevolt Series Operating and Service Guide
407
SCPI Programming Reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Truevolt Series and is the answer not in the manual?

Keysight Truevolt Series Specifications

General IconGeneral
BrandKeysight
ModelTruevolt Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals