7528%/(6+227,1* 9HUVD3XOVHPowerSuite6HULHV6HUYLFH0DQXDO
 5(9%
LED11 BLST GRN Indicates when the blast shield is inserted in its
slot. The ground return is by way of the chassis
ground straps on the laser deck.
LED12 FBRIN GRN Indicates when the fiber is fully inserted in the
connector.
LED13 FBRCON GRN Indicates when the fiber is inserted in the
connector.
LED14 CDRH GRN Indicates when the remote interlock connector is
inserted in the connector. If the interlock is used
in the operating room, this indicates that the
remote interlock is closed.
LED15 FSNC1 GRN Indicates that the normally closed contact in
footswitch #1 is closed.
LED16 FSNO1 GRN Indicates that the normally open contact in
footswitch #2 is closed.
LED17 FSNC2 GRN Indicates that the normally closed contact is
footswitch #1 is closed.
LED18 FSNO2 GRN Indicates that the normally open contact in
footswitch #2 is closed.
LED19 FSER2/FSDN2 RED/GRN If the LED is illuminating in the red, then
footswitch #1 is in an error state, both contacts
low, high, etc. If the LED is illuminating in the
green, then footswitch #1 is fully depressed and
debounced.
LED20 FSER2/FSDN2 RED/GRN If the LED is red, then footswitch #2 is in an error
state, both contacts are low or high. If the LED is
green, then footswitch #2 is fully depressed and
debounced.
LED21 NFIRE RED Indicates that the ‘No fire” bit from the DIO is
active and the HVPS is disabled.
LED22 SHUTCL/SHUTOP GRN/RED If the LED is green, then the safety shutter is
closed. If the LED is red, then the safety shutter
is open.