Maintenance...Messages
5-144
LTO
0xB Incompatible edger
configuration.
HI lens not allowed.
The job you have loaded calls
for High Index (HI) lens mate-
rial.
In the personalisation menus,
you have excluded this parame-
ter.
Therefore the edger cannot ac-
cept this job.
Confirm the message.
If you wish to load this job no-
netheless, change the personali-
sation of the pull-down menus.
Select High Index (HI) lens ma-
terial.
Load the job again.
0xC Incompatible edger
configuration.
Mineral lens not allowed.
The job you have loaded calls
for Mineral lens material.
In the personalisation menus,
you have excluded this parame-
ter.
Therefore the edger cannot ac-
cept this job.
Confirm the message.
If you wish to load this job no-
netheless, change the personali-
sation of the pull-down menus.
Select Mineral lens material.
Load the job again.
0xD Incompatible edger
configuration.
Organic lens not allowed.
The job you have loaded calls
for CR39 lens material.
In the personalisation menus,
you have excluded this parame-
ter.
Therefore the edger cannot ac-
cept this job.
Confirm the message.
If you wish to load this job no-
netheless, change the personali-
sation of the pull-down menus.
Select CR39 lens material.
Load the job again.
0xE Incompatible edger
configuration.
Polycarbonate lens not al-
lowed.
The job you have loaded calls
for Polycarbonate lens material.
In the personalisation menus,
you have excluded this parame-
ter.
Therefore the edger cannot ac-
cept this job.
Confirm the message.
If you wish to load this job no-
netheless, change the personali-
sation of the pull-down menus.
Select Polycarbonate lens mate-
rial.
Load the job again.
0xF Incompatible edger
configuration.
Bevel finishing not allowed.
The job you have loaded calls
for Bevel finishing.
In the personalisation menus,
you have excluded this parame-
ter.
Therefore the edger cannot ac-
cept this job.
Confirm the message.
If you wish to load this job no-
netheless, change the personali-
sation of the pull-down menus.
Select Bevel finishing.
Load the job again.
0x10 Incompatible edger
configuration.
Rimless finishing not al-
lowed.
The job you have loaded calls
for Rimless finishing.
In the personalisation menus,
you have excluded this parame-
ter.
Therefore the edger cannot ac-
cept this job.
Confirm the message.
If you wish to load this job no-
netheless, change the personali-
sation of the pull-down menus.
Select Rimless finishing.
Load the job again.
0x11 Incompatible edger
configuration.
Grooving finishing not
allowed.
The job you have loaded calls
for Grooving finishing.
In the personalisation menus,
you have excluded this parame-
ter.
Therefore the edger cannot ac-
cept this job.
Confirm the message.
If you wish to load this job no-
netheless, change the personali-
sation of the pull-down menus.
Select Grooving finishing.
Load the job again.
# Message title