AlgoMed Technical Reference Manual
Proprietary 101 of 110
Figure 67: Set units to Kgf
5. Verify that the displayed value is 0.00 Kgf
Figure 68: End of synchronization
In case the displayed value is not 0.00, press the ZERO button to initialize
the sensor again
6. Place the Calibration Verification Weight over the AlgoMed tip