EasyManua.ls Logo

Medoc AlgoMed - Algometer Calibration; Figure 69: Placing Weight; Figure 70: Weight Placed

Default Icon
110 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AlgoMed Technical Reference Manual
Proprietary 102 of 110
Figure 69: Placing weight
Figure 70: Weight placed
7. Verify that the weight measured by AlgoMed is according to the weight marked on the
calibration verification weight.
In case the displayed value is not according to the value marked on the
weight, repeat steps 1 7 until verification pass successfully.
14.3 Algometer Calibration
Algometer is supplied calibrated with NIST certificate.
It is recommended to calibrate the Algometer in a certified laboratory in case the calibration
verification fails.
For further information regarding calibration options please contact Medoc
Service.

Table of Contents