AlgoMed Technical Reference Manual
Proprietary 13 of 110
Warning - Connect the device, according to the local electrical
standards.
Disposal according to electronic scrap ordinance
Manual edition refers to current version of manufactured system
Medoc reserves the right to change specification without prior notice, in line
with policy of constant product improvements
1.7 Electromagnetic immunity
The AlgoMed is intended for use in the electromagnetic environment specified below. The
customer or the user of the AlgoMed should assure that it is used in such an environment.
Electromagnetic
environment –guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
8 kV contact
2, 4, 8, 15kV air
8 kV contact
2, 4, 8, 15kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-4
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Electrical fast
transient/burst
±2 kV for power
supply lines
±2 kV for power
supply lines
Mains power quality should
be that of a typical
commercial or hospital