EasyManua.ls Logo

Medrad Intego - EMC and Mobility Specifications

Medrad Intego
142 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
93
Appendix A - Cleaning and Maintenance
Calibrating the Pinch
Valves
1. Press the CONFIGURATION button on the Navigation Bar to access the Configuration
Screen.
2. Press the MAINTENANCE button and then press the Calibration tab.
3. Press the VALVE CALIBRATION button.
4. The system displays a message prompting the user to remove all tubing from the Pinch
Valves. Ensure that the tubing has been removed and Press OK.

Table of Contents

Related product manuals