27
2 - System Basics
3. Pull the Drive Override out until it is fully extended (approximately 15 cm (6 inches)).
4. Disengage the Caster Brakes.
WARNING: Crush hazard. Using the Drive Override disengages the Drive
System brakes. Ensure the Intego™ PET Infusion System is not on an incline or left
unattended when using the Drive Override and re-engage the Drive Override after
moving the system and engage the Caster Brakes.
5. Firmly grip both handles on the Operator Position of the Intego™ PET Infusion System.
6. Move the Intego™ PET Infusion System using firm, steady force. Avoid making jerking
movements or sudden starts and stops. Leave approximately 1 m (3 ft) per side to
complete turns.
7. Once the Intego™ PET Infusion System is in its new location, apply the Caster Brakes.
8. Push the Drive Override in to the normal operating position.
9. Unwrap the power cable from the Cable Storage and plug into an AC outlet.
Moving the Intego™ PET
Infusion System INT SYS
100
WARNING: Injury or damage to the system may occur if it is moved without the
Shielded Chamber Lid secured. Do not move system unless the Shielded Chamber
Lid is latched.
WARNING: Injury or damage may occur if the Intego™ PET Infusion System is
used on steep inclines. DO NOT use the system on a floor with an incline greater
than 10 degrees.
WARNING: Radiation exposure hazard. Do not move the Intego™ PET Infusion
System unless the Shielded Chamber Lid is latched.
CAUTION: DO NOT plug the Intego™ PET Infusion System power cable into an
extension cord or multi-outlet power strip. Only use the power cable supplied with
the system.
CAUTION: DO NOT hang clothing or other pieces of equipment from the
Intego™ PET Infusion System. Hanging clothes or equipment may impair the
movement of the system.
CAUTION: DO NOT move any portion of the Intego™ PET Infusion System by
force without disengaging the Caster Brakes or Brake Release.