Medrad™ Intego™
PET Infusion System
64
WARNING: Biological contamination can result from reusing the PAS or failure
to follow sterile technique. DO NOT reuse or attempt to resterilize. Properly discard
disposable items after use. If there is any possibility that contamination may have
occurred during set-up or use, disassemble and set-up a new sterile product.
WARNING: Patient infection may result from the use of non-sterile components.
Maintain sterility of all disposable components using aseptic techniques.
WARNING: Radiation exposure hazard. Higher residual radiation may occur
when using an extension tube or additional tubing between the catheter and the
PAS. The flush volume needs to be increased to account for additional tubing.
WARNING: Injury to patient may occur if system is moved without
disconnecting the PAS from the patient. Disconnect the PAS from the patient prior
to moving the Intego™ PET Infusion System.
Selecting the Flow Rate 1. The FLOW RATE button displays the currently configured flow rate. If this flow rate is not
the desired flow rate, press the button and select the desired flow rate from the displayed
list.
2. The selected flow rate will be displayed.