2 - System Basics
Moving the Intego™ PET
Infusion System INT SYS
200
WARNING: Injury or damage to the system may occur if it is moved without the
Shielded Chamber Lid secured. Do not move system unless the Shielded Chamber
Lid is latched.
WARNING: Injury or damage may occur if the Intego™ PET Infusion System is
used on steep inclines. Do not use the system on a floor with an incline greater than
10 degrees.
WARNING: Radiation exposure hazard. Do not move the Intego™ PET Infusion
System unless the Shielded Chamber Lid is latched.
WARNING: Trip hazard. Injury may occur. Do not operate Drive System if the power
cable is plugged into an AC outlet.
CAUTION: DO NOT plug the Intego™ PET Infusion System power cable into an
extension cord or multi-outlet power strip. Only use the power cable supplied with
the system.
CAUTION: DO NOT hang clothing or other pieces of equipment from the
Intego™ PET Infusion System. Hanging clothes or equipment may impair the
movement of the system.
CAUTION: DO NOT move any portion of the Intego™ PET Infusion System by
force without disengaging the Caster Brakes or Brake Release.
1. Disconnect the PAS from the patient.
WARNING: Injury to patient may occur if system is moved without
disconnecting the PAS from the patient. Disconnect the PAS from the patient prior
to moving the Intego™ PET Infusion System.
2. Unplug the power cable from the AC outlet and wrap the power cable on the Cable
Storage.
NOTE: If the Intego™ PET Infusion System is not placed in Standby and the power
cable is unplugged from the AC outlet, the Display will remain powered for up to
60 minutes after which the system will go into Standby.
NOTE: If the Intego™ PET Infusion System is in Standby and the power cable is not
plugged into AC power within 30 minutes, the Intego™ PET Infusion System will
completely Shutdown.
3. Disengage the Caster Brakes
NOTE: The Intego™ PET Infusion System cannot be moved using the Drive System if
the Device Lock feature is enabled and the Display is locked.
NOTE: The Intego™ PET Infusion System cannot be moved using the Drive System
unless the Shielded Chamber Lid is closed and the Shielded Chamber Lid Latch
is engaged.
NOTE: The Drive System cannot be engaged when priming the SAS or PAS,
performing a test injection, preparing an RP dose, or delivering an RP dose to
the patient or Waste Container.