75
6 - Vial Shield and SAS Removal
6 - Vial Shield and SAS Removal
Both the SAS and the Multi-Dose Vial may contain radiation even after the end-of-the-use
period. If possible, allow the used SAS and Multi-Dose Vial to decay to background levels (at
least 10 half-lives) before removing. If removal of the used SAS is required before decay to
background has been achieved, take radiation precautions per facility radiation safety
procedures.
NOTE: The Multi-Dose Vial and used SAS may be left within the Shielded Chamber
until the radiation level has decayed sufficiently for safe removal.
WARNING: Radiation exposure hazard.
18
F-FDG or
18
F-NaF emit gamma
radiation. Operator should exercise extreme caution when handling
radiopharmaceuticals. Doses may exceed maximum permissible dose if appropriate
precautions are not properly exercised. Follow all individual facility guidelines for the
handling and disposal of radioactive materials. DO NOT open the Shielded Chamber
without first taking radiation precautions per facility procedures.
WARNING: Puncture hazard. The SAS contains needles. To prevent injury, take
care when handling the SAS to control the needles. Dispose of the needles per
individual facility guidelines.
WARNING: Radiation exposure hazard. The Intego™ PET Infusion System may
contain radioactive materials. Do not reach under cart while the system contains
activity. Do not remove access panels while the system contains activity.
WARNING: Severe operator injury can occur if care is not taken when opening
or closing the lids over the Shielded Chamber or the Patient Administration Set
compartment. Be certain that all fingers are clear before opening or closing the
Shielded Chamber Lid or the PAS Compartment Cover.
1. Remove the Saline Tube from the Saline Pump and disconnect SAS from the Saline
Container. The Saline Container and/or the Saline Tube may leak when disconnected.
Take precautions to prevent spills.
2. Remove the Swabbable Valve and the Patient Tube from the Swabbable Valve Holder
and Air Detector.
3. Open the PAS Compartment Cover. Release the Shielded Chamber Lid Latch and slide
the Shielded Chamber Lid towards the front of the cart far enough to expose the two Pinch
Valves.
4. Remove the Waste Tube from the Waste Pinch Valve and the Patient Tube from the PAS
Pinch Valve. Press the button on top of the Pinch Valve and pull the SAS out.
5. Completely open the Shielded Chamber Lid and remove the RP Tube from the RP Pump.
6. To remove the SAS Needles from the Multi-Dose Vial, grasp the Needle Insertion Device’s
handle, lift straight up until it stops, then rotate counter-clockwise until it stops. Release
the handle and the Needle Insertion Device will remain in position.
7. Using the Vial Shield Carrying Handle, insert and attach the Vial Shield Access Cap by
pushing down and rotating clockwise. Ensure that the Access Cap is fully engaged with
the Vial Shield. Do not remove the Vial Shield at this time.
8. Lift the hinged Needle Insertion Device cover to access the Needle Cartridge holder.
Remove the Needle Cartridge from the Needle Insertion Device.