65
5 - Patient Infusion
Performing a Saline Test
Inject
NOTE: The Test Inject is optional.
NOTE: The Intego™ PET Infusion System will prevent a Saline Test Inject with less
than three minutes remaining on the battery meter.
1. On the Dosing screen, press the TEST INJECT button.
2. The Intego™ PET Infusion System displays a message asking the user to confirm that a
check for air has been completed. Inspect the PAS Tubing for air.
a. If the PAS contains air, press NO. Disconnect the PAS from the patient and return to
the "Priming the PAS" section of the manual.
b. If the PAS does not contain air, press YES. The Intego™ PET Infusion System will
infuse the patient with a small, preset volume of saline. See "System Configuration."
WARNING: Patient Injury could result from not following standard extravasation
minimizing techniques. To reduce the risks of extravasation conditions, follow all
standard extravasation risk minimizing techniques.