7 - Training Mode
7 - Training Mode
WARNING: Radiation exposure hazard. DO NOT install RP into the Intego™ PET
Infusion System while in Training Mode. If the system detects radiation in the Dose
Calibrator while operating in Training Mode, it will indicate an error and operation will
stop.
WARNING: Explosion hazard. DO NOT use the Intego™ PET Infusion System
when flammable gases are present. Patient injury could result from using the
system in the presence of flammable gases (such as anesthetics).
WARNING: While in Training Mode, the Intego™ PET Infusion System should
only be used for training purposes. Do not use the system to infuse fluid into a
patient.
NOTE: The Intego™ PET Infusion System will simulate normal functionality in Training
Mode. The user can perform all of the tasks detailed in the "Daily Setup and
Patient Infusion" section of this manual. Saline or other non-radioactive fluid
must be substituted for an RP to simulate an infusion.
NOTE: The Intego™ PET Infusion System cannot perform the Daily QC Constancy and
Accuracy Test, Dose Calibrator Check, Linearity check, or Dose Calibrator
Calibration while in Training Mode. MEDRAD recommends that the user
performs these tasks prior to using the system in Training Mode.
NOTE: In Training Mode, the pinch valve calibration is simulated and not stored when
the system is powered Off.
NOTE: The Intego™ PET Infusion System cannot simulate occlusions in Training
Mode.
1. Power on the Intego™ PET Infusion System.