Medrad™ Intego™
PET Infusion System
50
6. Press the REPLACE button in the Saline section.
7. Enter the volume of the Saline Container using the keypad.
NOTE: The minimum recommended Saline Container volume is 750 ml.
Installing the SAS NOTE: Always install a new SAS before installing a new Multi-Dose Vial.
NOTE: The Intego™ PET Infusion System is intended for use with
18
F-FDG and
18
F-
NaF. Intego™ PET Infusion System disposables intended for use with
18
F-FDG
may be used to deliver
18
F-FDG or
18
F-NaF. Please contact MEDRAD for more
information.
NOTE: Installing the SAS or installing the Multi-Dose Vial may be performed before or
after entering RP Assay and Saline Information.
NOTE: In the following steps, it will be helpful to refer to the diagrams of the Intego
Infusion System Components, Intego Shielded Chamber Components, and the
SAS Components.
WARNING: DO NOT use the SAS with more than one Multi-Dose Vial. Using the
SAS with more than one Multi-Dose Vial could result in injury to patient or operator.
1. Place the Saline Container onto the Saline Hook at the rear of the cart.
2. Open the PAS Compartment Cover.
3. Grasp the Shielded Chamber Lid Latch to release it, and slide the Shielded Chamber Lid
towards the front of the cart to gain access to the Shielded Chamber.
WARNING: Severe operator injury can occur if care is not taken when opening
or closing the Shielded Chamber Lid or the PAS Compartment Cover. Be certain
that all fingers are clear before opening or closing the Shielded Chamber Lid or the
PAS Compartment Cover.