4 - Daily Setup
4. The Vial Assay Information screen appears. Press the respective fields to enter the Lot
Number, Date, Reference Time, RP Type, Activity at Reference Time, and Volume. When
all data is entered, press the OK button.
WARNING: Radiation exposure hazard. Do not insert more than
27.75 GBq (750 mCi) into the Intego™ PET Infusion System.
NOTE: If the level of radioactivity in the Multi-Dose Vial exceeds 27.75 GBq (750 mCi),
the Intego™ PET Infusion System displays the message “Vial activity exceeds
safe shielding level. Immediately return vial to protective container.” If this
message appears, remove the Vial Shield from the system and return to hot lab
or other storage capable of providing adequate shielding for the contained
activity.
5. The System Preparation screen reappears with the usable activity in the Multi-Dose Vial
displayed within the vial symbol. The total Multi-Dose Vial volume and activity are
displayed to the right of the vial symbol.