Appendix C - Troubleshooting Tips
WARNING: In case of occlusion, do not disconnect disposables before
discarding dose, as liquid leakage can occur.
System Messages The following is a list of messages the Intego™ PET Infusion System could display and
possible corrections:
Critical Error Messages
NOTE: These Errors Require the System to be Restarted
Recoverable Errors / Messages
Code Error Messages Explanation
E1000 The measured concentration does not
match the expected value based on the
<Selected RP> assay information.
Confirm that the assay information was entered correctly.
Re-enter if needed.
Confirm that the assay information provided by the
pharmacist is correct. Re-enter if needed.
If the assay corrections do not correct the problem,
replace the Multi-Dose Vial.
If problem persists, recalibrate the Dose Calibrator.
If the problem continues, record the error message and
contact MEDRAD.
E1055 Activity detected while in Training Mode. Remove activity from the Dose Calibrator and restart the
Intego™ PET Infusion System in Training Mode.
If the problem continues, record the error message and
contact MEDRAD.
Any other
E code
N/A Restart the system.
If the message reappears, record the error message and
contact MEDRAD.
Error Messages Explanation
A dose cannot be prepared from this vial. Replace vial. No activity is available to extract.
Replace vial.
Activity concentration is less than the acceptable
minimum.
Concentration may have decayed below minimum.
Verify assay information was entered correctly and
reprime.
Verify SAS is installed properly and reprime.
Replace vial and reprime.
Activity concentration too high. Check assay information. Confirm that the assay information was entered
correctly. Re-enter if needed.
Confirm that the assay information provided by the
pharmacist is correct. Re-enter if needed.
If the assay was entered correctly, wait until the
concentration decays to 3.7 GBq/ml (100 mCi/ml).
Additional saline flush required. Press OK to flush to
waste.
A prior dose infusion was aborted or high residual
activity remains from priming the SAS.
It should not be necessary to re-prime the SAS, but if
problem persists, re-prime the SAS. Refer to the "If an
Error Occurs During or After Priming" section in the
"Troubleshooting" Appendix.
Air detected in SAS. Need to prime. Confirm that the SAS is correctly inserted into the Air
Detector.
Reprime the SAS. Refer to the "If an Error Occurs
During or After Priming" and the "Recovering from
Activation of the Air Detector" sections in the
"Troubleshooting" Appendix.