Medrad™ Intego™
PET Infusion System
112
Assay information incomplete. Enter all required data. • Enter all required information. Source lot number,
activity, and date/time are all required. Refer to "Daily
QC" section in the Clinical Use chapter and "Dose
Calibrator Calibration" section of this manual.
Calibration already in progress. • Wait until the measurement of the source is complete
before proceeding to the next step.
Check printer paper. • Check printer and replace paper as needed. Refer to
the "Replacing the Printer Paper" section of this
manual.
• Close the printer door.
Date/Time write failure. Please try again. • Re-enter the date/time. Refer to the "System
Configuration" section of the "System Basics" chapter.
Dose Calibrator is detecting high background activity.
Check for nearby sources.
• Check for nearby sources.
• Check for the calibration source in the Dose Calibrator.
Remove if still present.
• Check for nearby sources of radiation. If necessary,
remove the other source(s) or move the cart.
• Check for contamination in the room or in the system,
including possible radiopharmaceutical leak from the
SAS. Clean the system per site guidelines and replace
the SAS.
Dose Calibrator serial number mismatch. Recalibration
required.
• Recalibrate the Dose Calibrator. Refer to the "Dose
Calibrator Calibration" section of this manual.
Dose cannot be infused. Discard and prepare new dose. • Discard the current dose and prepare a new dose.
Error reading preferences, resetting system to defaults. • Re-enter system preferences on Configuration screen.
Infuse or discard current dose before preparing another. • A patient dose is already prepared. Either infuse or
discard the current dose before attempting to prepare
another dose or perform a test injection.
Invalid date/time entered. • Re-enter the date/time. Refer to the "System
Configuration" section of this manual.
Measured activity is higher than expected. Check assay
information and prime again.
• Confirm that the assay information was entered
correctly. Re-enter if needed. Refer to the "Entering
Multi-Dose Vial Assay Information" section of the
"Daily Use" chapter.
• Confirm that the assay information provided by the
pharmacist is correct. Re-enter if needed. Reprime.
Measured activity is low. Ensure vial and SAS are
properly installed. Replace if necessary.
• Confirm that the assay information was entered
correctly. Re-enter if needed. Refer to the "Entering
Multi-Dose Vial Assay Information" section of the
"Daily Use" chapter.
• Confirm that the assay information provided by the
pharmacist is correct. Re-enter if needed.
• Ensure vial and SAS are properly installed. Refer to
the "If an Error Occurs During or After Priming" section
in the "Troubleshooting" Appendix.
Operation will exceed Waste Container capacity. Replace
SAS.
• The Waste Container is full. Replace the SAS.