Appendix C - Troubleshooting Tips
Pinch Valve error: <Valve Error Details> See Operation
Manual.
Confirm that the SAS is properly installed into the
Pinch Valves. Refer to the "If an Error Occurs During
or After Priming" section in the "Troubleshooting"
Appendix.
Valve Error Details will be one of the following phrases:
waste valve failed.
patient valve failed.
both valves failed.
both valves open simultaneously.
tubing absent from waste valve.
tubing absent from patient valve.
tubing absent from both valves.
patient valve failure, tubing absent from
waste valve.
waste valve failed, tubing absent from
patient valve.
If necessary, restart the system. If the message
reappears, record the error message and contact
MEDRAD.
Prime PAS. The PAS has not been primed and the operation
cannot be completed. Prime the PAS before
proceeding.
Printer error occurred. Press the OK button and the system should correct the
issue.
If the message reappears, restart the system.
If the message persists, record the error message and
contact MEDRAD.
Residual activity is higher than the acceptable
background level. See Operation Manual.
High residual activity remains from priming the SAS.
Refer to the "If an Error Occurs During or After
Priming" section in the "Troubleshooting" Appendix.
Re-prime the SAS.
If re-priming does not resolve the issue, check for
contamination and replace the SAS.
Saline is low. Please replace. Replace the Saline Container. Refer to the "Daily
Setup" section.
Saline prime failed. Ensure SAS is properly installed.
replace if necessary.
Confirm that the SAS is correctly installed. Refer to the
"If an Error Occurs During or After Priming" section in
the "Troubleshooting" Appendix.
Replace the SAS if necessary.
SAS detected. Press Cancel to use or remove SAS and
press OK.
The system requires that the vial and SAS be removed
before the Multi-Dose Vial assay information can be
reset.
The SAS should only be used with a single Multi-Dose
Vial. Change the SAS each time a new Multi-Dose Vial
is installed. Refer to the "Installing the Multi-Dose Vial
and SAS" section of the "Daily Setup" chapter.
•Press CANCEL if the REMOVE button was pressed by
mistake.
<Selected RP> or saline information missing. Enter all
required data.
Enter all required information. Multi-Dose Vial Lot
Number, Assay (or Reference) Date and Time, Total
Activity, RP Volume, and Saline Volume are all
required. Refer to the "Entering Multi-Dose Vial Assay
Information" and "Entering Saline Data" sections of the
"Daily Setup" chapter.
<Selected RP> prime failed. Check SAS and vial
installation, then reprime.
Confirm that the SAS is correctly installed. Refer to the
"If an Error Occurs During or After Priming" section in
the "Troubleshooting" Appendix.
Replace the SAS if necessary.