103
Appendix B - Specifications
Electromagnetic
Compatibility (EMC)
US: The Intego™ PET Infusion System, although exempt per 47CFR15.103(c)(e), has been
verified to conform to the requirements of 47CFR15 as a Class B digital device.
EU: The Intego™ PET Infusion System has been verified to conform to the requirements of the
Medical Device Directive (93/42/EEC).
Mobility of INT SYS 200
Fluid to Patient
Flow rate 0.5 ml/s See note below.
1.0 ml/s See note below.
NOTE: Dose activity and fluid delivery accuracies are only guaranteed if MEDRAD Intego™
PET Infusion System disposables are used and if the patient connection contains no
restriction smaller than a 24 gauge, or larger, catheter, or 23 gauge or larger steel
needle. Use of smaller catheters or needles will affect flow rate and delivered volume
and may affect delivered dose activity.
Nominal Value Notes
PAS
Extended length 155 cm (60 in) Approximate length.
Shelf life 5 years ---
SAS
Waste Container volume 375 ml or 100 mCi ---
Shelf life 3 years ---
Nominal Value Notes
Speed, maximum forward,
high range
0.67 m/s (2.2 ft/s) ---
Speed, maximum forward,
low range
0.4 m/s (1.3 ft/s) ---
Speed, maximum reverse 0.4 m/s (1.3 ft/s) ---
Recharging time 8 h See note below.
Braking Automatic electric Brakes are applied by the
mobility controller when the
driver controls are released.
Manual Override --- Emergency use only. Pulling the
handle releases the drives,
allowing the system to be moved
manually.
Caster Brakes Manual User-applied to maintain
position.