Appendix D - Vials and Vial Shields
Appendix D - Vials and Vial
Shields
NOTE: Radiation Shielding performance is achieved by using the vial shield designed
by MEDRAD (or its equivalent).
WARNING: Patient injury may result from using vials with other than a 20 mm
diameter cap. Use of vials having other than a 20 mm diameter cap may create a
particulate hazard.
WARNING: No modification of this equipment is allowed.
WARNING: Biological contamination may occur if aseptic techniques are not
followed. Follow facility guidelines to disinfect the vial shield.
WARNING: Radiation Exposure Hazard. The Intego™ PET Infusion System Vial
Shields must be transported in a container that complies with applicable
transportation regulations.
CAUTION: If using an absorbent disc, place no more than one disc in the vial
shield before installing an RP vial. Additional absorbent discs may result in
improper fit and potential damage to the vial.
CAUTION: If using an absorbent disc, do not expose the absorbent disc to
cleaning solutions or other fluids. Discard and replace the absorbent disc if it has
been exposed to any fluid. Failure to replace the absorbent disc that has been
exposed to fluids may result in improper fit and potential damage to the vial.
CAUTION: If using an absorbent disc, discard and replace the absorbent disc
after each use. Failure to replace the absorbent disc may result in improper fit and
damage to the vial.
CAUTION: Do not place labels, tags, or other materials on the exterior of the vial
shield surface. Application of labels, tags, or other materials may cause the vial
shield to fit incorrectly in needle insertion device and result in improper fit and
functioning of the system.
CAUTION: Ensure that the needle insertion guide is clean and free of debris.
Debris may cause the vial shield to fit incorrectly in needle insertion device and result
in improper fit and functioning of the system.