Medrad™ Intego™
PET Infusion System
114
Source not detected. During calibration, the source was not inserted into the
Dose Calibrator. Insert the appropriate source to
continue with calibration. Refer to the "Dose Calibrator
Calibration" section of this manual.
The date and time must be set. Set the date/time. Refer to the "System Configuration"
section of the "System Basics" chapter.
The process has been interrupted and must be restarted. The STOP button has been pressed. To resume
operation, do one of the following:
During SAS and PAS Priming
Start the priming sequence again by pressing the
PRIME button.
During Dose extraction
Press the DISCARD button and then press the
PREPARE button to extract a new dose.
During Test Injection
Press the RESUME button to resume the test
injection, or
Press the STOP button to stop the test injection.
During patient infusion
Press the RESUME button to resume the infusion, or
Press the DISCARD button to discard the dose.
Process paused. During Test Injection
Press the RESUME button to resume the test
injection, or
Press the STOP button to stop the test injection.
During patient infusion
Press the RESUME button to resume the infusion, or
Press the DISCARD button to discard the dose.
The supplier activity is too high to perform the requested
operation.
Activity exceeds 25.9 GBq (700 mCi).
Confirm that the assay information was entered
correctly. Re-enter if needed.
Confirm that the assay information provided by the
pharmacist is correct. Re-enter if needed.
If the assay was entered correctly, wait until the activity
decays to 25.9 GBq (700 mCi).
This action is not allowed in Training Mode. Dose Calibrator calibration and the Linearity check can
not be performed while the system is in Training Mode.
Restart the system in Clinical Mode to perform a Dose
Calibrator calibration or a Linearity check.
Unable to detect printer. Press OK to retry. Press the OK button and the system should correct the
issue.
If the message reappears, restart the system.
If the message persists, record the error message and
contact MEDRAD.
Vial activity exceeds safe shielding level. Immediately
return vial to protective container.
Activity exceeds 27.75 GBq (750 mCi), which is the
maximum safe shielding level for the system.
The Vial Shield should be removed from the system
until the total activity in the Multi-Dose Vial decays to
27.75 GBq (750 mCi).
Waste activity will exceed shielding limit. Replace SAS. Activity in the Waste Container exceeds
3.7 GBq (100 mCi).
Replace SAS before continuing operations.
Residual activity exceeds 2% of original dose. Press OK
to flush to waste.
High residual activity remains after infusion.
Refer to the "PAS Recovery" section in the
"Troubleshooting" Appendix.
PAS Occlusion Detected. WARNING PAS is pressurized
and may leak if disconnected.
If the infusion is stopped prematurely, do not
disconnect disposables before discarding dose, as
liquid leakage can occur.
Reference Appendix C - Trouble Shooting Tips PAS
Occlusion Recovery.