Appendix C - Troubleshooting Tips
Appendix C - Troubleshooting Tips
Recovering from Priming
Issues
If an error occurs during or after priming, the following should be checked for proper
installation.
WARNING: Radiation exposure hazard.
18
F-FDG or
18
F-NaF emit gamma
radiation. Operator should exercise extreme caution when handling
radiopharmaceuticals. Doses may exceed maximum permissible dose if appropriate
precautions are not properly exercised. Follow all individual facility guidelines for the
handling and disposal of radioactive materials. DO NOT open the Shielded Chamber
without first taking radiation precautions per facility procedures.
CAUTION: Incorrect installation of the SAS may result in improper functioning
of the Intego™ PET Infusion System.
Ensure the Saline Container is sufficiently spiked.
Confirm the tubing is placed correctly in the Saline Pump.
Confirm the tubing is oriented on the Saline Pump rollers properly.
Ensure the tubing is not pinched in the Saline Pump door.
Ensure the Saline Disconnect Luer is against the Saline Pump.
Confirm that the Needle Cartridge Clip is inserted with proper orientation into the Needle
Insertion Device.
Confirm the collar on the tubing is positioned properly in the RP Pump.
Confirm the tubing is inserted into the Pinch Valves correctly.
Confirm the tubing is fully inserted into the Air Detector.
Confirm the RP Pump is closed and the tubing is in the pump.
Confirm that all tubing is laying flat and not being pinched by the lid.
Recovering from
Activation of Air Detector
During Priming
1. Confirm that the SAS is properly inserted into the Air Detector. If necessary, slide the
Shielded Chamber Lid open to expose the Air Detector and properly insert the SAS.
WARNING: Radiation exposure hazard.
18
F-FDG or
18
F-NaF emit gamma
radiation. Operator should exercise extreme caution when handling
radiopharmaceuticals. Doses may exceed maximum permissible dose if appropriate
precautions are not properly exercised. Follow all individual facility guidelines for the
handling and disposal of radioactive materials. DO NOT open the Shielded Chamber
without first taking radiation precautions per facility procedures.
2. Re-prime the system.
3. The Display displays a message stating that the SAS has already been primed. Press the
YES button to continue.
During a Saline Test Inject
1. Confirm that the SAS is properly inserted into the Air Detector. If necessary, slide the
Shielded Chamber Lid open to expose the Air Detector and properly insert the SAS.