26
Medrad™ Intego™
PET Infusion System
NOTE: The Intego™ PET Infusion System must be powered On or in Standby to use
the Drive System. If the system is powered Off, the Drive System will not
operate.
NOTE: The battery must be charged for the Drive System to function. The blue Status
Indicator will be illuminated when the Drive System is available for use. If the
blue Status Indicator is blinking, plug the Intego™ PET Infusion System power
cable into an AC outlet to charge the battery.
4. Choose either slow or fast on the Drive Speed Switch.
5. Firmly grip both of the handles on the Intego™ PET Infusion System at the Operator
Position. Squeeze the Drive Engage Switch.
6. To move forward, use the thumb of the right hand to slowly rotate the Drive Controller
forward while keeping the Drive Engage Switch pressed. To move backward, use the
thumb of the right hand to slowly rotate the Drive Controller backward while keeping the
Drive Engage Switch pressed.
NOTE: If the blue Status Indicator begins blinking while moving the Intego™ PET
Infusion System, there are approximately 3 minutes remaining before the Drive
System will stop functioning. When the blue Status Indicator begins blinking,
move the system to an AC outlet and plug in the power cable.
7. Move the Intego™ PET Infusion System by carefully pushing the handles to the left and
right to steer. The speed can be varied using the Drive Controller. The further the Drive
Controller is rotated from the center position, the faster the system will move.
CAUTION: Avoid making fast turns or sudden starts and stops.
NOTE: Use forward motion to cross thresholds while moving the Intego™ PET Infusion
System.
8. Once the Intego™ PET Infusion System is in its new location, release the Drive Engage
Switch or the Drive Controller to stop.
9. Apply the Caster Brakes.
10. Unwrap the power cable from the Cable Storage and plug into an AC outlet.
NOTE: To ensure that the battery is fully charged and has the maximum capacity, keep
the Intego™ PET Infusion System connected to AC power at all times except
when the system is being moved.
Moving the System Using the Drive Override (For Emergency Use Only)
NOTE: If the Drive System is not available, the Intego™ PET Infusion System can be
moved by using the Drive Override.
1. Disconnect PAS from patient.
2. Unplug the power cable from the AC Outlet. Wrap the power cable on the Cable Storage.