Medrad™ Intego™
PET Infusion System
70
NOTE: The Intego™ PET Infusion System may detect an occluded line to the patient
during an infusion. Appendix C, PAS Recovery, provides instructions for
handling occlusions detected during an infusion.
WARNING: In case of occlusion, do not disconnect disposables before
discarding dose, as liquid leakage can occur.
Infusion Completion
After an infusion completes, the Intego™ PET Infusion System displays a summary screen that
shows the amount of RP delivered (both volume and activity), total fluid volume delivered, and
any patient information provided.
1. Press the PRINT button to print this summary.
2. Press the OK button to update the infusion history and close the summary window.
NOTE: The Intego™ PET Infusion System should not be put in Standby before
pressing the OK button.
NOTE: The printed label does not have all the information included on the Display.
NOTE: After the infusion completes, disconnect the PAS from the patient and Intego™
PET Infusion System. Although PAS residuals are negligible for infusions
without occlusions, the PAS should be disposed of using the facility procedure
for disposal of radioactive material.
NOTE: After pressing the OK button, the main Dosing screen appears.