Medrad™ Intego™
PET Infusion System
100
Environmental
Electrical
Nominal Value Notes
Non-Operating Conditions
Temperature, maximum ambient 60º C (140º F) ---
Temperature, minimum ambient -20º C (-4º F) ---
Humidity, maximum 100% R.H. Non-condensing
Humidity, minimum 5% R.H. Non-condensing
Pressure, maximum barometric 48 kPa (7 psi) ---
Pressure, minimum barometric 110 kPa (16 psi) ---
Vial Shield Temperature,
maximum ambient
70º C (158º F) ---
Vial Shield Temperature,
minimum ambient
-40º C (-40º F) ---
Operating Conditions
Temperature, maximum ambient 40º C (104º F) ---
Temperature, minimum ambient 10º C (50º F) ---
Humidity, maximum 90% R.H. ---
Humidity, minimum 20% R.H. ---
Pressure, maximum barometric 69 kPa (10 psi) ---
Pressure, minimum barometric 110 kPa (16 psi) ---
Nominal Value Notes
Mains Power Connection
Mains supply voltage, maximum 240 VAC ---
Mains supply voltage, minimum 100 VAC ---
Mains supply frequency,
maximum
60 Hz ---
Mains supply frequency,
minimum
50 Hz ---
Power consumption, maximum,
INT SYS 200
300 VA ---
Power consumption, maximum,
INT SYS 100
250 VA ---
Safety
Electrical leakage current,
earth and chassis
< 100 µA See note below.
Electrical leakage current,
patient connection
< 10 µA See note below.