Medrad™ Intego™
PET Infusion System
1. Disconnect the PAS from the patient.
WARNING: Injury to patient may occur if system is moved without
disconnecting the PAS from the patient. Disconnect the PAS from the patient prior
to moving the Intego™ PET Infusion System.
NOTE: If the Intego™ PET Infusion System is not placed in Standby and the power
cable is unplugged from the AC outlet, the Display will remain powered for three
minutes. If the power cable is not plugged into the AC outlet within three
minutes, the system will go into Standby.
NOTE: If the Intego™ PET Infusion System is in Standby and the power cable is not
plugged into AC power within 30 minutes, the system will completely power Off.
2. Unplug the power cable from the AC outlet and wrap the power cable on the Cable
Storage.
3. Disengage the Caster Brakes
4. Firmly grip both handles on the Operator Position of the Intego™ PET Infusion System.
Squeeze the Brake Release on each handle to disengage the brakes.
NOTE: The Brake Release on each handle must be squeezed simultaneously to
disengage the brakes.
NOTE: To engage the brakes, stop squeezing either Brake Release.
5. Move the Intego™ PET Infusion System using firm, steady force. Avoid making jerking
movements or sudden starts and stops. Leave approximately 1 m (3 ft) per side to
complete turns.
6. Once the Intego™ PET Infusion System is in its new location, stop squeezing the Brake
Release to stop.
7. Apply the Caster Brakes.
8. Unwrap the power cable from the Cable Storage and plug into an AC outlet.