Medrad™ Intego™
PET Infusion System
86
3. Press the NEW MEASUREMENT button to begin a new Linearity check.
4. Enter the activity.
NOTE: The Intego™ PET Infusion System automatically measures the activity level. If
the clinician enters a value for activity, that value will be retained for future
reference and reporting.
a. Press the Activity field.
b. Using the keypad, enter the activity and reference time from the assay information.
Press the ENTER button when complete.
5. Using the Calibration Source Holder, place vial of
18
F into the Dose Calibrator and close
the Shielded Chamber Lid.