EasyManua.ls Logo

MICRO-EPSILON CS02 - General Settings

MICRO-EPSILON CS02
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 56
Ethernet Interface
capaNCDT 6200
All measuring values are transmitted as Int32. The measuring value resolution is 24 bits i.e only the most low-
order 24 bits of integer number are used. Hexadecimal range: 0 ... FFFFFF
16
. Exception is the mathematics
function, because the result can also be greater than 24 bits.
Scaling of measuring values:
Measurement in µm = * Measring range in µm
Digital value (Int)
0xFFFFFF
Example: Measuring range sensor CS2 = 2000 µm; digital value = 7FFFFF
16
Measurement = 999.99 µm
By default, the measuring values are continuously output with the set data rate via the data port.
However, there is also a trigger mode, with can be used to get the individual measuring values, see 6.4.2.
6.3 Settings
Operating modes:
- Continuous transmission with fixed data frequency
- Trigger mode (recall hardware trigger input or individual measuring values, see 5.4.
Data rate:
It is possible to adjust different data rates between 2.5 Sa/s and 3.9 kSa/s. The data rate applies to all chan-
nels.
Filter/Measuring value averaging:
The following filters are selectable:
- Moving average
- Arithmetic average (only each n
th
value will be output)
- Median
- Dynamic Noise Rejection
The setting for the averaging applies to all channels.

Table of Contents

Related product manuals