Page 62
Ethernet Interface
capaNCDT 6200
6.4.4 Filter, Averaging Type (AVT)
Mode of measurement averaging
AVT
Command $AVTn<CR>
Response $AVTnOK<CRLF>
Index
n = 0: No averaging (Default setting)
n = 1: Moving average
n = 2: Arithmetic average (outputs only each nth measuring value)
n = 3: Median
n = 4: Dynamic noise rejection
? = Request averaging type
Request averaging type
Command
$AVT?<CR>
Response
$AVT?nOK<CRLF>
6.4.5 Filter, Averaging Number (AVN)
Number of measuring values used to calculate the average (adjustable from 2 … 8)
AVN
Command $AVNn<CR>
Response $AVNnOK<CRLF>
Index n = 2 ... 8
? = Request averaging number
Request averaging number
Command $AVN?<CR>
Response $AVN?nOK<CRLF>