5-29 
6.    If the BIS and NMT modules are configured, perform the following patient auxiliary current 
tests between each electrode and the others by turns: 
  Normal polarity; 
  Reverse polarity; 
  Normal polarity with open neutral; 
  Reverse polarity with open neutral; 
  Normal polarity with open earth; 
  Reverse polarity with open earth; 
7.    Verify that the maximum leakage current does not exceed 100 μA (0.1 mA) in the first two 
tests. While for the last two tests, verify that the maximum leakage current does not exceed 
500 μA (0.5 mA). 
 
  Ensure the safety analyzer is authorized by certificate organizations (UL, CSA, 
AAMI, etc.). Follow the instructions of the analyzer manufacturer. Fluke ESA620 
is used as an example. 
Electrical safety inspection form 
Location:  Technician: 
Equipment:  Control Number: 
Manufacturer:  Model:  SN: 
Measurement Equipment/SN:  Date of Calibration: 
Inspection and Testing  Pass/Fail  Limit 
1  Auxiliary mains output voltage     
2  Earth resistance  Ω    Max 0.1 Ω 
3 
Earth leakage 
current 
Normal condition  ____μA    Max:   
Normal condition: 500 
μA 
Single fault condition: 
1000 μA 
Single fault 
condition 
____μA   
4 
Patient 
leakage 
current 
Normal condition  ____μA    Max:   
Normal condition: 100 
μA 
Single fault condition: 
500 μA 
Single fault 
condition 
____μA   
5  Mains on applied part  ____μA    Max: 5000 μA 
6 
Patient 
auxiliary 
current 
Normal condition  ____μA    Max:   
Normal condition: 100 
μA 
Single fault condition: 
500 μA 
Single fault condition  ____μA