EasyManuals Logo

Mindray A8 Service Manual

Mindray A8
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #162 background imageLoading...
Page #162 background image
5-30
In routine maintenance, all the test items included in the electrical safety inspection form shall be
performed. The following table specifies test items to be performed after the equipment is
repaired with the main unit disassembled.
When none of the power board, transformer, and patient
circuit board is repaired or replaced
Test items: 1 and 2
When the power board or transformer is repaired or replaced Test items: 1, 2, and 3
When the patient circuit board is repaired or replaced Test items: 1, 2, 4, 5, and 6
When all of the power board, transformer, and patient circuit
board are repaired or replaced
Test items: 1, 2, 3, 4, 5, and 6

Table of Contents

Other manuals for Mindray A8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A8 and is the answer not in the manual?

Mindray A8 Specifications

General IconGeneral
BrandMindray
ModelA8
CategoryMedical Equipment
LanguageEnglish

Related product manuals