6-48
3. Select Begin to access the screen shown below.
NOTE
The real-time fresh gas flow is displayed on the Calibrate Total Flow Sensor
screen. Normally, only N2O and total flow values are greater than 0, and N2O
and total flow values ascend with the increase of calibration progress. Air and
O2 values should be 0 or small all the time.