7-18
Message
Priority
Cause
Solution
2. If the gas supply on the pipeline is
sufficient, but "Air Supply Failure" is
still reported, replace the air supply
inlet assembly (refer to 11.1.8).
3. If the cylinder pressure gauge
indicates that the cylinder pressure is
sufficient, but the status screen
indicates that the pressure is lower
than the normal range, replace the air
cylinder yoke (refer to 11.1.2).
N2O Supply
Failure
Med
N2O supply pressure is
low.
1. Check the connection of the N2O
supply and ensure that the gas supply
is normal. The gas supply pressure on
the AIR pipeline should be within the
range of 280- 600kPa (40-87PSI),
and the cylinder pressure should be
within the range of
4.2-6.0MPa(600-870PSI). Check the
pressures displayed on the pressure
gauge and status screen.
2. If the gas supply on the pipeline is
sufficient, but "N2O Supply Failure"
is still reported, replace the N2O
supply inlet assembly (refer to 8.1.8).
3. If the cylinder pressure gauge
indicates that the cylinder pressure is
sufficient, but the status screen
indicates that the pressure is lower
than the normal range, replace the
N2O cylinder yoke (refer to 11.1.2).
7.3.4.1 EFCS Error Codes
Failure
Code
Cause FRU Solution
0x00000001
Control module
watchdog selftest
error
EFCS control
board
Restart the machine. If the problem
persists, replace the EFCS control board
(refer to 11.3.1).
0x00000002
Control module
EEPROM selftest
error
EFCS control
board
Restart the machine. If the problem
persists, replace the EFCS control board
(refer to 11.3.1).
0x00000004
Protection module
selftest error
EFCS control
board
Restart the machine. If the problem
persists, replace the EFCS control board
(refer to 11.3.1).
0x00000008
O2 proportional
valve selftest error
EFCS assembly
Restart the machine. If the problem
persists, replace the EFCS assembly
(refer to 11.1.4).
0x00000010
Balance gas
proportional valve
error
EFCS assembly
Restart the machine. If the problem
persists, replace the EFCS assembly
(refer to 11.1.4).
0x00000020 O2 limb leakage EFCS assembly
Restart the machine. If the problem
persists, replace the EFCS assembly

Table of Contents

Other manuals for Mindray A8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A8 and is the answer not in the manual?

Mindray A8 Specifications

General IconGeneral
BrandMindray
ModelA8
CategoryMedical Equipment
LanguageEnglish

Related product manuals